Repeated Timing Jitter Measurement Summary
======================================================================

Bias Voltage: 16.00 V
Trigger Voltage: -0.430 V
Number of measurements: 10
Acquisition time per measurement: 10.0 s

Individual Measurements:
----------------------------------------------------------------------
Run	Peak_1_FWHM(ps)	Peak_2_FWHM(ps)	Countrate(cps)
1	36.28	39.67	452352.9
2	37.07	39.49	444718.8
3	36.66	39.54	436372.6
4	36.82	39.61	421447.9
5	36.79	39.62	410962.7
6	36.63	39.73	400441.7
7	36.92	38.95	385588.9
8	36.09	39.28	376222.1
9	36.30	38.93	361116.9
10	38.60	37.73	349990.7

======================================================================
Statistics:
----------------------------------------------------------------------
Peak 1 Average FWHM: 36.81  0.70 ps
Peak 2 Average FWHM: 39.26  0.61 ps
Average Countrate: 403921.5  35396.3 cps