Repeated Timing Jitter Measurement Summary
======================================================================

Bias Voltage: 16.00 V
Trigger Voltage: -0.430 V
Number of measurements: 10
Acquisition time per measurement: 10.0 s

Individual Measurements:
----------------------------------------------------------------------
Run	Peak_1_FWHM(ps)	Peak_2_FWHM(ps)	Peak_3_FWHM(ps)	Countrate(cps)
1	32.50	35.70	44.79	667355.8
2	32.70	35.01	41.31	665876.1
3	32.69	35.14	45.13	669531.5
4	32.90	35.29	44.80	658142.4
5	32.77	35.00	42.71	659830.6
6	32.76	35.34	44.65	659358.9
7	32.90	35.29	45.87	666380.0
8	32.74	35.00	46.41	658373.2
9	32.46	35.27	41.99	659775.6
10	32.68	35.25	40.15	657822.0

======================================================================
Statistics:
----------------------------------------------------------------------
Peak 1 Average FWHM: 32.71  0.14 ps
Peak 2 Average FWHM: 35.23  0.21 ps
Peak 3 Average FWHM: 43.78  2.10 ps
Average Countrate: 662244.6  4486.3 cps